The FALCON 400 G2 covers micro- Vickers , Vickers, Knoop , and Brinell tests in a single instrument – ​​with test forces from 1 gf to 62,5 kgf. This allows it to replace two machines in many laboratories: one for micro-forces and one for low-force and macro-forces.

The test force is generated via a multi-load cell in a closed control loop, entirely without weights. The tolerance is less than 0,5 percent across all force ranges.

the essentials in brief

  • Testing procedure: Micro-Vickers, Vickers, Knoop, Brinell
  • Inspection personnel: 1 gf to 62,5 kgf – configurable in steps and retrofittable
  • Test force tolerance: less than 0,5 percent across all strength ranges
  • Power generation: Multiple load cells in a closed control loop, without weights
  • Revolver head: Motorized, 6 positions for 2 intruders and 4 lenses, with collision detection and automatic retraction
  • Auto Focus: Automatic focusing when changing lenses
  • Objective: 5x, 10x, 20x, 50x, 100x with large working distance
  • OK: electronic-digital 15x, analog optionally or additionally
  • Z-axis: two motorized ball screw axes, dynamic electronic handwheel
  • XY table: manual, 100 x 100 mm, travel range 25 x 25 mm
  • Workpiece holder: 145 mm height x 170 mm depth
  • Steering: 6,5-inch color touchscreen with I-TOUCH workflow control
Front view of the INNOVATEST FALCON 400 G2

A device for micro, small and macro loads

The test force range of 1 gf to 62,5 kgf is unusually wide for this class of equipment. In practice, this means that the same machine can measure the case hardening depth on a surface layer with HV0,1 and a weld seam with HV10 – without changing equipment, without second calibration, and without a second workstation.

You don't order the FALCON 400 G2 with the full range, but rather with the force range you actually need. Six configurations are available. If your testing requirements change, the range can be expanded later – without replacing the machine. This makes the machine an investment that grows with your laboratory.

Would you like even more information?

Click here & download PDF!

Vickers, Knoop and Brinell scales and norms

The 6-round revolver accepts up to two indenters. Brinell-type bullets in 1 mm, 2,5 mm, and 5 mm sizes are available.

Process scale range Standards
Vickers HV0,001 to HV60 ISO 6507, ASTM E384, ASTM E92, JIS B 7725
Button HK0,001 to HK5 ISO 4545, ASTM E92, JIS Z 2251
Brinell HBW1/1 to HBW5/62,5 ISO 6506, ASTM E10, JIS Z 2243

The conversion between the scales – also to Rockwell , Leeb or tensile strength – is carried out according to ASTM E140, ISO 18265 and GB/T 1172.

For accredited testing laboratories according to ISO/IEC 17025, regular indirect testing with traceable hardness reference plates is mandatory. Direct and indirect calibration, including certification according to ISO and ASTM, NADCAP, and ILAC-MRA, is available upon request. We handle initial calibration, recurring tests, and documentation.

Dual Z-axis: Separate positioning and inspection

Most hardness testers move the workpiece and the test force along the same axis. The FALCON 400 G2 separates these two movements onto two independent motorized ball screw axes.

The first axis positions the workpiece quickly, at up to 30 mm per second. The second axis applies the test force with high resolution. What would be a compromise between speed and sensitivity with a single axis is here distributed across two drives.

In addition, there is an electronic handwheel whose ratio depends on the lens used: large steps per revolution at low magnification, and correspondingly finer steps at 50x or 100x with shallow depth of field. This dynamic handwheel is only available on the 400 G2 within the FALCON series.

Motorized turret head with autofocus

The turret offers six positions for up to two indenters and four lenses. Changing lenses is part of the automated inspection cycle – no reconfiguration is required. During lens changes, the workpiece surface remains automatically in focus, eliminating the need for manual refocusing.

If the turret detects a collision with an indenter or lens, the system automatically retracts the workpiece. Neither the machine nor the sample is damaged.

I-TOUCH Workflow Control

Operation is via a 6,5-inch color touchscreen with I-TOUCH firmware. The display sits on its own stand and can be positioned and tilted for right- or left-handed users – easily readable whether standing or sitting.

Test programs can be created, saved, and recalled; limit values ​​can be defined as pass/no-go; and results can be converted to other hardness scales. Statistics—maximum, minimum, mean, range, standard deviation—are recorded in real time after each measurement. Data is exported via USB.

Optional: IMPRESSIONS evaluation software

The IMPRESSIONS package includes an 11 MP camera, a 27-inch touchscreen, and a system controller. No additional PC is required. The software recognizes impressions even on scratched surfaces, automatically adjusts brightness and contrast, and performs standardized data analysis.

In addition, there are application modules for CHD, SHD and NHD, weld seam testing according to ISO 9015, edge decarburization according to ISO 898-1 , 2D and 3D hardness mapping and Kic crack measurement.

Technical details FALCON 400 G2

feature Value
test procedure Micro-Vickers, Vickers, Knoop, Brinell
Testing staff area 1 gf to 62,5 kgf, configurable in steps and retrofittable
Power generation Multiple load cells, closed-loop control system with force feedback, without weights
Test force tolerance < 0,5% across all test force ranges
Holding time 10-second standard, freely adjustable
turret head Motorized, 6 positions: 2 intruders, 4 lenses, collision detection with automatic retraction
Autofocus Automatic focusing when changing lenses
Objectives 5x, 10x, 20x, 50x, 100x (large working distance)
Okular electronic-digital 15x · analog 15x optionally or additionally
camera 11 MP via the optional IMPRESSIONS camera and software system
Z axis two motorized ball screw axes, dynamic electronic handwheel, positioning up to 30 mm/s
XY table Manual, 100 x 100 mm, travel 25 x 25 mm, scale 0,01 mm; max. 60 kg up to 2 kgf, max. 100 kg over 2 kgf
Workpiece holder 145 mm (H) x 170 mm (D)
Advertising 6,5-inch color touchscreen, tiltable, can be mounted on the left or right · 27-inch LCD with IMPRESSIONS package
Display resolution 0,1 HV / HK · 0,5 HB
revaluation Rockwell, Rockwell surface, Vickers, Brinell, Knoop, Leeb, tensile strength according to ASTM E140, ISO 18265, GB/T 1172
Data output USB
Standards ISO 6507, ISO 4545, ISO 6506, ASTM E384, ASTM E92, ASTM E10, JIS B 7725, JIS Z 2251, JIS Z 2243

FALCON 300, 400, 450 or 500 – which device is right for you?

The FALCON series differs primarily in its test force range and degree of automation:

Model Inspection personnel Typical use
FALCON 300 G2 5 gf – 2 kgf Pure microhardness testing, Vickers and Knoop. Manual cross table, finely adjustable Z-axis via side handwheel.
FALCON 400 G2 1 gf – 62,5 kgf Micro, low-power, and macro capabilities in one device, plus Brinell resistance. Force range upgradeable, dual motorized Z-axis with dynamic handwheel.
FALCON 450 G2 200 gf – 62,5 kgf Stand-alone device for macro and micro Vickers, Knoop, and Brinell measurements. Z-axis adjustment via handwheel, optional video measurement system.
FALCON 500 G2 0,1 gf – 62,5 kgf Multi-load cell, fully automatic and freely configurable. Motorized Z-axis with ball bearing spindle.

Are you unsure which model is right for your testing needs? Send us your materials and test specifications and we'll recommend the perfect device!

Interested? Contact us now!

Ask us now without obligation.

E-mail Phone